ECD-4-nano


ECD-4-nano is a high-resolution electrochemical dilatometer with a capacitive parallel plate sensor offering sub-5 nm resolution, ideal for monitoring electrode or full cell stack thickness changes during electrochemical cycles.

 

1. Redesigned Test Cell
  • Corrosion-resistant cell bottom for use with a wide range of electrolytes
  • New One-Seal concept significantly improves tightness compared to ECD-3
  • Enables stable long-term operation

 

 2. Integrated into PAT System

  • Can be inserted directly into PAT-Tester-x or docking station such as PAT-Clamp-1
  • Space-saving and fast setup
  • Integrated PAT-Button ensures automatic recognition in EL-Software

 

3. Use Cases

  • Expansion of individual electrode (3-electrode setup with ring-shaped reference electrode)
  • Expansion of full cell stack (2-electrode setup)
  • For aprotic electrolytes

 

4. Additional Key Features
  • Capacitive displacement sensor (range 250 μm, resolution ≤ 5 nm) arrangements
  • Additional gas pressure sensor (0–3 bar) and temperature sensor (-20 to 80°C)
  • Cableless connection via PAT socket with electronic cell tag (PAT-Button)

 

 

Length: 67 mm 
Width: 67 mm
Height: 129 mm
Weight: 2 kg
Sample diameter: 10 mm
Glass T-Frit (Separator): Diameter 12.5 mm / 10 mm, Thickness 3.5 mm
Separator diameter (full cell mode): max. 10 mm
Working (upper) electrode diameter: max. 10 mm
Counter (lower) electrode diameter: max. 10 mm
Gas pressure sensor: 0 to 3 bar abs.
Temperature sensor: -20 to 80° C
Chemical compatibility: Aprotic organic electrolytes
Cell electrolyte volume: Half cell mode: approx. 0.2 ml, Full cell mode: approx. 0.03 ml
Operational temperature range (cell and sensor): -20 to 80° C
Operational temperature range (conditioning electronics and data logger): 0 to 40 °C
Load on test specimen: approx. 1 Newton
Displacement sensor type: Capacitive
Displacement range: 250 μm
Displacement resolution: ≤ 5 nm